Intraoperative Mains Oxygen Supply Failure - Guidelines for Crises in Anaesthesia
Complete failure of wall or pendant oxygen supply. Failure may be theatre-specific, zone-specific or hospital-wide
❶ Inform the theatre team of the problem, including theatre coordinator (Box A).
❷ Call for help, but be aware failure may be widespread.
❸ Switch to cylinder oxygen supply.
❹ Check remaining cylinder content (Box B):
• Ensure at least one further cylinder always available.
❺ Prolong duration of cylinder oxygen supply:
• Use circle system.
• Use lowest possible fresh gas flow.
• Use lowest possible FiO2.
• Check FiO2 alarms.
❻ Minimise auxiliary gas usage:
• Some ventilator systems are gas-driven. If so, switching to manual ventilation will prolong oxygen cylinder supply.
❼ Evaluate moving to nearby area if oxygen supply there preserved.
❽ Continue to monitor and ensure adequacy of inspired oxygen and volatile agent.
❾ Workflow:
• Do not start any new case unless clinical priority absolutely requires it.
• Expedite conclusion of current case if clinically appropriate.
• Consider recovering the patient in theatre if recovery also affected (Box C).
❿ Disconnect from failed outlets and do not re-connect until safe to do so (Box D).
By Association of Anaesthetists @ https://twitter.com/AAGBI
Quick Reference Handbook - Guidelines for crises in anaesthesia